Reliable protection against air embolisms
LIFEBRIDGE®’s patented 7 stage safety system protects reliably against air embolisms by completely eliminating air bubbles, again, fully automatically. Replaceable “patient modules” – including encapsulated components that come into blood contact directly – are disposed of by the hospital after use. In stationery use, the system operates with 110/220 VAC mains power. In mobile 24 V battery operation, it will run for up to 2 ½ hours under average load.

Modular concept
The LIFEBRIDGE®’s modular design facilitates its handling before, during, and after its service. Each basic unit (beige) comprises at least one red control module and one grey patient module. After each use, the patient module is detached and discarded (see drawing) an the control module is shipped to LIFEBRIDGE Medizintechnik for refurbishing.

Menu driven start-up
The display guides the user through the LIFEBRIDGE®’s start-up in five steps. This guarantees safe operations even in time-critical emergencies. A rotary knob and a touch screen are used to select the clearly structured menu’s options.
![]() |
![]() |
|
Control unit and base station specifications
| General information | |
| Preparation time | A few minutes |
| Air embolism protection | 7 stage air elimination system |
| Display | TFT touch screen |
| Dimensions | 24.0 in x 17.7 in x 14.6 in (H/H/W)* |
| Weight | |
| Control module | approx. 40 pounds (18 kg) |
| Base station | approx. 40 pounds (18 kg) |
| Power supply | |
| External power supply unit | Input: 220-240 VAC, 50/60Hz, 450VA, output: 31 VDC, 14.5 A |
| Base station battery life time | approx. 120 minutes |
| System battery life time | approx. 30 minutes |
| Displayed values | |
| Flow | 0.0 to 9.9 l/min (accuracy: ± 0.1 l/min) |
| Pressure | -400 to +400 mmHg (accuracy: ± 10 mmHg) |
| Rotational speed | 0 to 4000 rpm (accuracy: ± 50 rpm) |
* without base station
Patient module specifications
| Oxygenator | |
| Casing material | Polycarbonate |
| Heat exchanger fibre | Polyurethane |
| Gas exchanger fibre | Polypropylene, microporous |
| Gas exchanger surface | 1.35 m² |
| Blood flow rate | 1 to 7 l/min |
| Deaeration volume (static) | 190 ml |
| Heat exchanger surface | 0.08 m² |
| Water connections | ½ in Hansen couplings |
| Filter | |
| Casing material | Polycarbonate |
| Filter material | Polyester |
| Pore size | 40 μm |
| Maximum blood flow | 8 l/min |
| Centrifugal pump | |
| Maximum output pressure | 900 mmHg (0.9 kPa) |
| Tubes and connectors | |
| Blood filled tubes | Silicone |
| Connector material | Polycarbonate |
Product approvals
30th December 2007 – CE certification
The LIFEBRIDGE®
is the world’s first CE certified emergency cardiopulmonary support system. To achieve this, the complex safety-related requirements for the system were identified in close coordination with Munich based TÜV Süd Product Services, one of the leading certification organisations in this special field of medical technology, and were then successfully implemented within the LIFEBRIDGE®
product development.
17th October 2009 – FDA clearance
LIFEBRIDGE Medizintechnik has been awarded clearance by the U.S. Food and Drug Administration (FDA) for a portable heart-lung machine, a first for any manufacturer worldwide. Since its launch in early 2008, more than 40 hospitals have adopted the LIFEBRIDGE®
in Germany alone. After its successful FDA clearance it is also gaining ground in the U.S.
T +49 8636 69603-0
F +49 8636 69603-25